Zirc #4 Illuminator Mouth Mirror (10pk)
Next image
Zoom image

Zirc #4 Illuminator Mouth Mirror (10pk)


Not rated yet
In stock (9 items available)

Zirc offers the most brilliant and quality mouth mirror lens on the market today. The Crystal HD® mirror lens is 40% brighter than Rhodium and reduces eye strain and fatigue.

 

Photos courtesy of: Dr. Dimitris Tsanaktsidis | www.dental-blog.gr

  • $46.00
Qty:  
  • Weight
    2.2 lbs
  • SKU
    50Z301

 

  • Highest-Grade Front-Surface-Rhodium-Coated lens
  • Reinforced angle provides extra strength when used as a retractor
  • Mirror is flush with plastic to prevent plaque build-up
  • Autoclavable up to 136°C/275°F
  • Contains Antimicrobial Protection
  • Washer-disinfect
  • 4 to 6 weeks delivery

 

Please follow the instructions for use from the manufacturer of your cleaning and sterilization equipment. Hereinafter called “see manufacturer

information”.

1. Restriction of Reprocessing: Frequent reprocessing has marginal consequences for the Mouth Mirrors. Normally the end of durability of the product will be

determined by visual degradation.

2. Preparation Before Cleaning: Place the Mouth Mirrors into your instrument holder (ex. Cassette).

3. Cleaning: There are three (3) cleaning methods that can be used to clean your Mouth Mirrors and Handles.

*Manual: Clean the Mouth Mirrors manually by washing with a non-corrosive, low sudsing neutral detergent.

*Ultrasonic: “see manufacturer information”

*Automated Washer: “see manufacturer information”

After one of the above cleaning methods have taken place. Rinse Mouth Mirrors thoroughly.

4. Drying: Dry Mirrors thoroughly before sterilization. IMPORTANT: Failure to dry the mirrors thoroughly may cause marks to be burned into the glass surface

or may lead to corrosion.

5. Sterilization: Steam sterilize for at least 5 minutes at 270° F/134° C or 20 minutes at 250° F/121° C. Do not heat above 275° F/136° C. To avoid corrosion the

steam has to be free of ingredients. Recommended critical values of ingredients for table water and steam condensate are determined according to EN 285.

Please note: After sterilization in the Autoclave, please pay attention to the drying period INSIDE the Autoclave, do not take them out before drying period ends.

DO NOT “dry-heat” sterilize.

20Z340LT-3 Rev. E 01/09